-D width: The width of the first negative pulse in the 50% amplitude points.
+Duty: +Duty Cycle, defined as +Width/Period.
-Duty:-Duty Cycle, defined as -Width/Period.
Delay A→B : The delay between the two channels at the rising edge.
Delay A→B : The delay between the two channels at the falling edge.
Screen Duty: Defines as (the width of the positive pulse)/(Entire period)
Phase A→B : Phase difference calculated according to " Delay A→B "
and the period of source A, expressed in degree. The calculation formula
is as shown below:
Phase A→B = (Delay A→B
÷ Period of source A) × 360°
Phase A→B : Phase difference calculated according to " Delay A→B "
and the period of source A, expressed in degree. The calculation formula
is as shown below:
Phase A→B = (Delay A→B
÷ Period of source A) × 360°
Note for the following delay measurements:
Source A and Source B can be set in the automatic measurement function menu.
FRR: Time between Source A first rising edge and Source B first rising edge.
FRF: Time between Source A first rising edge and Source B first falling edge.
FFR: Time between Source A first falling edge and Source B first rising edge.
FFF: Time between Source A first falling edge and Source B first falling edge.
LRR: Time between Source A first rising edge and Source B last rising edge.
LRF: Time between Source A first rising edge and Source B last falling edge.
LFR: Time between Source A first falling edge and Source B last rising edge.
LFF: Time between Source A first falling edge and Source B last falling edge.
Other measurements
+PulseCount : The number of positive pulses that rise above the
mid reference crossing in the waveform.
-PulseCount
: The number of negative pulses that fall below the
mid reference crossing in the waveform.
RiseEdgeCnt
: The number of positive transitions from the low
reference value to the high reference value in the waveform.
FallEdgeCnt
: The number of negative transitions from the high
reference value to the low reference value in the waveform.