EasyManua.ls Logo

Philips HeartStart XL - Secondary Method

Philips HeartStart XL
270 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
M4735A Service Manual 4-31
Opening the Case
4
Secondary Method
a. Using two clip leads, connect one clip lead to each end of one of the
Disarm Resistors.
b. Touch the clip leads directly to the spade terminals on the Power PCA
where the Defibrillator Capacitor is connected. Maintain this
connection for at least 5 seconds.
TIP: Another way to make this connection is to clip the leads to test
probe leads, then use the test probes to touch the spade terminals.
c. Repeat steps a and b using the other Disarm Resistor. The
defibrillator capacitor is now discharged, even if one of the resistors
is damaged.
Figure 4-18 Discharging the Defibrillator Capacitor
High Voltage Discharge Tool
Touch Here First
Then Touch Here
Disarm Resistors
Defibrillator Capacitor
Foam Base

Table of Contents

Other manuals for Philips HeartStart XL

Related product manuals