7-1
7
7 Specifications
Overview
This section provides:
z Specifications for the M4735A,
z Symbol Definitions for symbols appearing on the M4735A,
z Safety related information, and
z Electromagnetic compatibility information.
Specifications
Defibrillator
Waveform: Biphasic Truncated Exponential. Waveform parameters adjusted
as a function of patient impedance.
For details of waveforms delivered, see the “Waveforms” section of this
chapter.
Shock Delivery: Via multifunction defib electrode pads, or paddles.
Delivered Energy Accuracy: See Table 7-1.