M4735A Service Manual 7-7
Specifications
7
Events can be marked with a Mark symbol and, if configured for drug annota-
tion, the following labels can be added: Epinephrine (Adrenaline in U.K. and
Australia), Atropine, Lidocaine, or Other.
The Summary key on the front panel is used to print the internal Event Sum-
mary.
Data Card Event Summary:
The Data Card stores continuous ECG waveforms and events.
General
Dimensions: 19.0 cm (H) x 37.6 cm (W) x 34.6 cm (L)
7.5" x 14.8" x 13.7"
Weight: Standard Configuration weighs 6.5 kg (14.3 lbs.) including battery,
full roll of paper, and external paddles.
Environmental
Temperature: 0
o
to 55
o
C operating, -20
o
to 70
o
C storage.
z Thermal paper may darken above 55
o
C.
z Charging the battery at temperatures above 35oC may degrade battery
life.
z Storing the battery for extended periods at temperatures above 40oC
will reduce battery capacity and degrade battery life.
Humidity: Up to 95% Relative Humidity.
z Printer may jam if paper is wet.
z Printer may be damaged if wet paper is allowed to dry while in con-
tact with the printhead elements.
Altitude: Operating: up to 15,000 ft. Storage: up to 15,000 ft.
Shock: Philips Medical Systems Section 760, Class B1 Drop Test.
Vibration: Philips Medical Systems Corp. Section 759 Class B1 Vibration.
Water Resistance: Meets IEC 601-2-4 (44.3),
IEC 529 Level IPX0 (non-protected).
Safety: Meets EN 60601-1
EMC: Meets EN 60601-1-2.
Other Considerations: Equipment not suitable for use in the presence of a
flammable anesthetic mixture with air, oxygen, or nitrous oxide.
Mode of Operation: Continuous.
AC Line Powered: 100-240 VAC, 50/60 Hz, .4A max (unit off),
1.5A max (while charging defibrillator).
Battery Powered: 12 V Rechargeable, SLA.