M4735A Service Manual 2-13
The Language Support Tool
2
4. Select the desired test or function.
To select a test, press the and softkeys to move up or down the
list until the desired test is highlighted. Then press the softkey to
start that test.
5. Exit Diagnostic Mode.
To exit Diagnostic Mode, turn the unit off.
The Language Support Tool
The Language Support Tool allows field service personnel to perform several
tasks:1) to set the language of the Control PCA of the defibrillator; 2) to
enable the SpO
2
option; 3) to program in the serial number. These tasks need
to be performed under the following circumstances:
z The Control PCA has been replaced.
The Control PCA contains all the operating software, configured
for the installed hardware. It also contains the unit’s serial number,
which was assigned and programmed during manufacturing.
The new Control PCA must be programmed to recognize the
hardware installed in this unit, to contain that unit’s
serial number, and to set the language of the unit.
z The unit has received an upgrade adding the SpO
2
capability.
The added hardware will not be automatically recognized. The
Control PCA must be programmed to recognize the new hardware
installed.
z The unit’s software is being updated.
Software upgrades are periodically released to resolve customer
issues. The Language Support tool enables you to load new software
onto the unit.
Using the Language Support Tool
1. Save the configuration to a Data Card.
a. Make sure the unit is powered off, and have either a fresh battery
installed or the AC power cord plugged in.
b. Follow the instructions in "Configuration Mode" on page 2-11 to save
the configuration to a Data Card.
ENTER