Test and Inspection Matrix
2-4 Performance Verification and Safety Tests
Test and Inspection Matrix
Table 2-1 summarizes performance verification tests and inspections for the
M4735A; including test name, test or inspection to perform, expected test
results, and data to record.
Table 2-1 Performance Verification and Safety Tests
Test Group Name
Test or Inspection to
Perform
Expected Test Results
Data to Record
x = p (pass) or f(fail)
Visual Inspection
(V)
Inspect the unit, accesso-
ries, cables, etc.for signs
of wear, damage, corro-
sion, or missing items, as
described on page 2-16.
If no unusual damage, no
corrosion, no missing items:
x=p.
then Visual Inspection
passes
V:x
Example V:p
Functional
Checks (F)
In normal Operating Mode,
perform the following func-
tional checks:
If all functions respond as
expected: x=p
F:x
Example F:p
• ECG (page 2-18). • Waveform clear on dis-
play; HR correct on display;
HR alarm works. Leads off,
pads off indicators as
expected.
• Shock Advisory (page 2-
19).
• Shock Advised only when
appropriate
• Synchronized Cardiover-
sion (page 2-20).
• Shock delivered with cor-
rect timing (<
60msec)
• SpO
2
(page 2-21). • 95% -100%
then Functional Check
passes
Extended Self
Test (X)
In Diagnostic Mode, run
the Extended Self Test
(page 2-25). Includes Data
Card Test and time/date
check.
If “Pass” reported on all tests
applicable to the device con-
figuration and options: x=p.
then Extended Self test
passes
X:x
Example X:p