Internal Assemblies - Top Case
4-40 Removal and Replacement
Shield Plate
The sections below describe how to remove and replace the metal Shield Plate
located under the Control PCA. This procedure is often needed when
removing and replacing other assemblies.
Preparation
Disconnect and remove the Control PCA.
See "Control PCA" on page 4-36.
Removal
a. Remove the five screws around the edges of the shield plate. See Fig-
ure 4-22 on page 4-40.
b. Lift the plate straight up.
Figure 4-22 Removing the Shield Plate
Keyscan PCA
Tab
Slot
SlotTab
Shield Plate