M4735A Service Manual 4-41
Internal Assemblies - Top Case
4
Replacement
a. Lower the plate into place. Be sure to engage the tabs on the Keyscan
PCA into the mating slots in the Shield Plate.
b. Replace the five screws.
TIP: Screw installation is easier if all screws are started in their holes
before any one of them is tightened.