General Problems
3-18 Troubleshooting
Extended Self Test Failures
Should the unit report a FAIL in the Extended Self Test, resolve it using the
solutions below.
NOTE Before replacing any parts, check to see if the cables and flex circuits are
properly connected. See "Servicing Notes" on page 4-1.
Table 3-6 Extended Self Test Failures
Failure Possible Cause Suggested Solution
RAM
ROM
System
CODEC
IRDA
Timebase
Failure on Control PCA. Replace Control PCA.
System FAIL 10
Lithium backup battery failure. Replace lithium battery.
Data Card
Data Card full, or incompatible, or
defective.
Replace Data Card with a new,
empty M3510A.
Failure on Control PCA. Replace Control PCA.
Pacer
Defib
Failure on Power PCA. Replace Power PCA.
Failure on Control PCA. Replace Control PCA.
FE
Failure on Control PCA. Replace Control PCA.
Failure on Power PCA. Replace Power PCA.
Failure on Parameter PCA. Replace Parameter PCA.
SpO2
Failure on SpO
2
PCA. Replace SpO
2
PCA.
Failure on Control PCA. Replace Control PCA.