Specifications
7-4 Specifications
Heart Rate Alarms: Configurable pairs of low and high heart rate alarm
limits: 30 to 100, 60 to 140, 90 to 160, and 120 to 200 bpm.
Hands Free Defibrillation Patient Cable Length: 7 ft. (2.13 m).
ECG Cable Length: 12 ft. (3.7 m).
Common Mode Rejection: Greater than 90 dB measured per AAMI standard
for cardiac monitors (EC 13).
ECG Size: 2.5, 5, 10, 20, 40 mm/mV.
Frequency Response:
AC Line Filter: 60 Hz/50Hz.
Pads ECG for Display: Monitor (.15-40 Hz) or EMS (1-30 Hz).
Pads ECG for Printer: Monitor (.15-40 Hz) or EMS 1-30 Hz).
Leads ECG for Display: Monitor (.15-40 Hz) or EMS (1-30 Hz).
Leads ECG for Printer: Diagnostic (.05-150 Hz) or EMS (1-30 Hz) or
Monitor (.15-40 Hz).
Patient Isolation (defibrillation proof):
ECG: Type CF
SpO
2
: Type CF
Defib: Type BF - External paddles/pads
Type CF - Internal paddles
Display
Size: 115 mm x 86 mm.
Type: EL - Electroluminescent.
Resolution: 320 x 240 pixels.
Sweep Speed: 25 mm/s nominal (stationary trace; sweeping erase bar).
Viewing Time: 4 seconds.
Battery
Type: 2 Ah, 12V, rechargeable, Sealed Lead Acid (SLA).
Dimensions: 2.4" (H) x 0.94" (W) x 7.2" (D).
61.7 mm (H) x 23.9 mm (W) x 182 mm (D).
Weight: 1.4 lb. (0.65 kg).
Charge Time:
z Approximately 3 hours to 90% charge, indicated by LED on front
panel.
z Approximately 15 hours total to 100%.