System Functional Descriptions
6-16 Theory of Operation
Lithium Backup Battery
The Backup Battery (located on the Control PCA) provides standby power to
maintain data during times when the main battery is either absent or
discharged and no external power is supplied. The data maintained includes
the user's configuration choices, and the system time and date.
Data Card
The Data Card allows the capture of key information such as ECG waveform,
shock advisories, charging, and delivering a shock. The Data Card is read by
the Code Runner Web system for post-event analysis.
CAUTION The Data Card must be inserted or removed only when the unit is powered
off. A Data Card inserted while the power is on will not be recognized by the
unit. If a Card is removed while the power is on, its stored data may be
corrupted and rendered unreadable.