Loading...
External Assemblies
4-10 Removal and Replacement
Battery Cover
The instructions that follow describe how to remove and replace the battery
cover.
Preparation
1. Remove the battery.
2. Remove the paddles from their retainers and unplug them from the
unit.
3. Turn the unit over.
Turn the unit upside down with the handle facing you. The battery cover
is at left side of the bottom surface.
Removal
1. Remove the battery cover.
Remove the 7 screws as shown in Figure 4-3.
TIP: The screws are size-coded. Remove only the T10 screws.
2. Lift the battery cover up.
Figure 4-3 Removing the Battery Cover
Replacement
1. Check the Battery PCA and the Battery Eject Spring.
a. Ensure the lower edge of the Battery PCA is seated in its slot in the
case