Test Equipment
2-10 Performance Verification and Safety Tests
Test Equipment
Table 2-2 lists the equipment needed to perform the Performance Verification
tests, and provides specifications for commercially available analyzers and
simulators. Test equipment is called out within each test procedure when
needed. In addition, A 50 ohm test load is available (M1781A or M3725A).
Table 2-2 Equipment List
Equipment/Test Specifications
ECG Simulator
Calibrated Leads
ECG simulator
Amplitude accuracy ±2%
Rate accuracy ±2%
Calibrated Paddles
ECG simulator
Amplitude accuracy ±2%
Rate accuracy ±2%
Defibrillator Analyzer
Waveform compatibility Meets all specs below using biphasic
truncated exponential waveform.
Load resistance: 50 ±1% (non-inductive)
Maximum energy: 200 joules
Maximum voltage: 2500 V
Maximum current: 50 A
Measurement accuracy:
20 joules: ±2% of reading
< 20 joules: ±0.4 joules
Cardioversion measurement range: –150 to +150 ms
Pacer tester
Load impedance: 400
Current
measurement accuracy
10 mA–50 mA: <±2 mA
50 mA–200 mA: <±4%
Rate
measurement accuracy
30–180 ppm: <±0.5%
Waveform duration accuracy:
30–180 ppm: ±1ms