M4735A Service Manual 2-9
Test and Inspection Matrix
2
External Paddles/Pads If readings are as expected:
Source
(Normal Condition) - ggg
<
100 uA
Source (Single Fault
Condition) - hhh
<
500 uA
With Mains on applied part
(Single Fault Condition) - iii
<
5000 uA
then External Paddles/Pads
Safety test passes
Internal Paddles If readings are as expected:
Source
(Normal Condition) - jj
<
10 uA
Source
(Single Fault Condition) -
kk
<
50 uA
With Mains on applied part
(Single Fault Condition) - lll
<
100 uA
then Internal Paddles/Pads
Safety test passes
Note: When recording test results, separate results within a test by slashes; separate tests by a semicolon (;); and
use no empty spaces. For example:
V:x;F:x;X:x;U:x;E:x;P:x;
DA:x;DB:x;D:x;
S1:P/aaa/bbbb;S3:P/aa/bb/cc/dd/ee/ff/ggg/hhh/iii/jj/kk/lll
V:p;F:p;X:p;U:p;E:p;P:p;
DA:p;DB:p;D:p;
S1:P/125/800; S3:P/9/49/49/10/50/50/100/499/4750/9/49/83
Table 2-1 Performance Verification and Safety Tests (Continued)
Test Group Name
Test or Inspection to
Perform
Expected Test Results
Data to Record
x = p (pass) or f(fail)

Table of Contents

Other manuals for Philips HeartStart XL

Related product manuals