Electromagnetic Compatibility
7-20 Specifications
Restrictions for Use
Artifact on the ECG caused by electromagnetic interference should be evalu-
ated by a physician or physician authorized personnel to determine if it will
negatively impact patient diagnosis or treatment.
Immunity Level
The EMC standards state that manufacturers of patient-coupled equipment
must specify immunity levels for their systems. It is recognized that the
M4735A defibrillator/monitor is designed to receive and amplify low level
signals in the same bandwidth as the interference.
Immunity is defined in the standard as the ability of a system to perform with-
out degradation in the presence of an electromagnetic disturbance. Degrada-
tion in ECG quality is a qualitative assessment which can be subjective.
Caution should, therefore, be taken in comparing immunity levels of different
devices. The criteria used for degradation is not specified by the standard and
may vary with the manufacturer.
NOTE For additional information about compliance with the EMC standards, visit
our Medical Systems website at www.medical.philips.com/cms and follow
the link to Regulatory.