M4735A Service Manual 4-51
Internal Assemblies - Top Case
4
After Repair
After this repair is complete, perform the following steps.
1. Replace and reconnect the Keyscan PCA.
See "Keyscan PCA" on page 4-42.
2. Replace the shield plate.
See "Shield Plate" on page 4-40.
3. Replace and reconnect the Control PCA.
See "Control PCA" on page 4-36.
4. Reassemble the case.
See "Closing the Case" on page 4-99.
5. Restore the customer’s configuration.
Enter Configuration Mode and reload the customers configuration from
the data card, or reconfigure the unit, as needed. See "Configuration
Mode" on page 2-11.
6. Test performance.
Conduct Performance Verification Testing as described in "Performance
Verification and Safety Tests" starting on page 2-1.