M4735A Service Manual 4-55
Internal Assemblies - Top Case
4
Figure 4-28 Removing the Display
After Repair
After repairs are complete, perform the following steps.
1. Replace the Bezel on the case.
See "Bezel Assembly" on page 4-48.
2. Replace and reconnect the Keyscan PCA.
See "Keyscan PCA" on page 4-42.
3. Replace the shield plate.
See "Shield Plate" on page 4-40.
4. Replace and reconnect the Control PCA.
See "Control PCA" on page 4-36.
5. Reassemble the case.
See "Closing the Case" on page 4-99.
6. Restore the customer’s configuration.
Enter Configuration Mode and reload the customers configuration from
the data card, or reconfigure the unit, as needed. See "Configuration
Mode" on page 2-11.
7. Test performance.
Conduct Performance Verification Testing as described in "Performance
Verification and Safety Tests" starting on page 2-1.
Display Assembly Display Ribbon Cables Energy Select Switch
Bezel Assembly
(Shown removed
from case