M4735A Service Manual 4-63
Internal Assemblies - Top Case
4
Figure 4-32 Removing the SpO
2
Connector
2.
Connect the SpO
2
Connector.
a. Connect the flex circuit from the SpO
2
connector to the SpO
2
PCA.
Line it up carefully, avoiding excessive twisting. There is no latch -
just push straight in. See Figure 4-30 on page 4-59.
After Repair
After repairs are complete, perform the following steps.
1. Reassemble the case.
See "Closing the Case" on page 4-99.
2. Restore the customer’s configuration.
Enter Configuration Mode and reload the customers configuration from
the data card, or reconfigure the unit, as needed. See "Configuration
Mode" on page 2-11.
3. Test performance.
Conduct Performance Verification Testing as described in "Performance
Verification and Safety Tests" starting on page 2-1.