Internal Assemblies - Top Case
4-68 Removal and Replacement
2. Replace the ECG Out (Sync) Connector.
a. Replace the ECG Out Connector in its hole in the case. Align one of
the flat sides against the ribs in the case.
NOTE One corner of the ECG Out connector is rounded. Keep that away from the
ribs, so the ribs can prevent the connector from rotating in either direction.
b. Replace the washer and nut on the connector, and tighten the nut.
c. Connect the 3-pin connector to the Parameter PCA. Be sure it latches
in place.
After Repair
After repairs are complete, perform the following steps.
1. Reassemble the case.
See "Closing the Case" on page 4-99.
2. Restore the customer’s configuration.
Enter Configuration Mode and reload the customer’s configuration from
the data card, or reconfigure the unit, as needed. See "Configuration
Mode" on page 2-11.
3. Test performance.
Conduct Performance Verification Testing as described in "Performance
Verification and Safety Tests" starting on page 2-1.