EasyManua.ls Logo

Philips HeartStart XL - Page 88

Philips HeartStart XL
270 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operational Problems
3-24 Troubleshooting
Charges, but disarms when press
SHOCK or paddles buttons.
Patient impedance sensed as too
high or too low during energy deliv-
ery due to:
Pads/paddles losing contact with
patient.
Pads/paddles failure.
Pads cable failure.
Replace paddles, or pads and pads
cable.
Power PCA failure. Replace Power PCA.
Control PCA failure. Replace Control PCA.
Charges, but disarms spontane-
ously.
1. Unit sensed
Pads Off or Cable
Off
due to:
Pads/paddles losing contact with
patient.
Pads/paddles failure.
Pads cable failure.
Power PCA failure.
Control PCA failure.
Keyscan PCA failure.
2.
DISARM key failure (intermittent).
1. Replace paddles, or pads and
pads cable.
2. Diagnose further as follows:
In Diagnostic Mode, run Controls
Test and test
DISARM key.
Key tests OK: Replace 1) Power
PCA 2) Control PCA.
Key not OK: Replace 1) Keyscan
PCA 2) Bezel Assembly 3) Control
PCA.
Charges slowly - about 4-5 sec.
instead of 2-3 sec.
Battery too old or not fully charged. Replace with new, fully charged
battery. Also run Battery Capacity
Test on suspect battery (see
page 2-47).
Unit senses unsupported battery
due to:
Unsupported battery in use.
Battery PCA failure.
Power PCA failure.
Control PCA failure.
Use only supported battery
(M3516A/M5516A).
Diagnose further as follows:
In Diagnostic Mode, start Battery
Capacity Test (see page 2-47).
Unsupported battery” on
display: Stop test. Replace
1) Battery PCA 2) Power PCA
3) Control PCA.
No message on display: Stop test.
Replace 1) Power PCA 2) Control
PCA.
Doesn’t deliver correct energy into
Defibrillator Analyzer or delivers no
energy at all. (Should also get mes-
sage - No shock delivered or Defib
failure
.)
Control PCA failure. Replace Control PCA.
Power PCA failure. Replace Power PCA.
Table 3-9 Operational Problems - Defibrillation and Synchronized Cardioversion (Continued)
Message or Symptom Possible Causes Suggested solution

Table of Contents

Other manuals for Philips HeartStart XL

Related product manuals