Defibrillator
Charge
Check
Shock
hazard.
When
charged
and
discharged
during
this
TCP,
the
defibrillator
discharges
up
to
360J
of
electrical
energy
through
the
defibrillator
electrodes.
Unless
discharged
safely
as
described
in
this
TCP,
this
electrical
energy
may
cause
injury
or
death.
Do
not
attempt
to
perform
this
procedure
unless
you
are
thoroughly
familiar
with
the
operation
of
the
defibrillator.
Shock
hazard.
Terminals
and
wires
carrying
high
voltages
up
to
+11.6kV
are
exposed
with
the
instrument
cases
removed.
Do
not
touch
internal
circuitry
except
as
indicated
in
these
procedures.
Possible
component
damage.
PCBs
in
this
instrument
contain
Static
Sensitive
Devices
(SSDs).
Use
the
Special
Handling
Procedures
for
Static
Sensitive
Devices
described
in
Section
4
of
your
Service
Manual
to
prevent
build
up
of
static
charge.
Refer
to
Table
5-1
in
your
Service
Manual
to
identify
PCB
assemblies
containing
SSDs.
.
Remove
the
Front
Panel
(see
the
Disassembly
Procedures
in
Section
4
of
your
Service
Manual).
Press
the
ON
pushbutton
on
the
Front
Panel.
.
Use
the
DMM
to
monitor
the
voltage
at
the
J3
pin
25
as
shown
in
NO
TAG
(connect
the
DMM
ground
to
J3
pin
7).
.
With
the
instrument
power
off,
install
the
Shock Advisory
Adapter
or
Defibrillation
Adapter.
.
Remove
the
battery
door
on
the
bottom
of the
instrument
and
disconnect
the
battery.
Connect
the
DC
power
supply
to
the
battery
connector
inputs.
Connect
ground
to
pin
1
or
2
(black
wire)
and
power
to
pin
3
4
(red
wire).
.
Turn
the
instrument
power
on
and
initially
set
the
power
supply
output
to
+18.00
+0.05Vdc.
6.
Charge
the
defibrillator
to
360J.
7.
Gradually
reduce
the
power
supply
voltage
toward
the
upper
Low
Battery Limit
detection
threshold
of
+15.22Vdc.
.
With
the
defibrillator
cables
attached
to
the
test
load
connectors,
press
both
adapter
discharge
pushbuttons.
.
Repeat
steps
6.
through
8.
above,
lowering
the
power
supply
voltage
toward
the
lower
Low
Battery
Limit
detection
threshold
of
+14.78Vdc
until
the
LOW
BATTERY:
CONNECT
AC
POWER
message
is
displayed.
Confirm
that
the
triggering
voltage
is
within
the
Low
Battery
Limits
shown
in
the
Low
Battery
Thresholds
Table.
All
Rights
Reserved
©
October
1994
Physio-Control
Corporation