LIFEPAK
9
defibrillator/monitor
PIP-Rate
Alarms/
1.
Press
HR
ALARM.
Confirm
that
the
HR
ALARM
LED
is
on
and
150/40
Alarm-Activated
is
displayed
in
reverse
video.
Record
2.
While
150/40
remains
displayed
in
reverse
video,
press
HR
ALARM
twice
and
confirm
that
120/60
and
160/90
display
consecutively.
3.
Press
HR
ALARM
to
display
150/40
in
reverse
video.
Confirm
that
after
about
2
seconds
150/40
displays
in
normal
video.
4.
|
Set
the
ECG
simulator
rate
to
greater
than
150bpm.
Confirm
that
the
alarm
sounds,
the
150
flashes,
and
the
recorder
runs
for
approximately
16
seconds.
5.
Confirm
that
the
heart
rate
printed
on
the
printout
is
151bpm
or
higher.
6.
Press
HR
ALARM
and
confirm
that the
alarm
stops.
7.
Decrease
the
ECG
simulator
rate
to
30bpm.
Press
HR
ALARM.
Confirm
that
the
alarm
sounds,
the
40
flashes,
and
the
recorder
runs
for
approximately
16
seconds.
8.
Press
HR
ALARM
and
confirm
that
the
alarm
stops.
PIP-Recorder
1.
Set
the
ECG
simulator
to
60bpm.
j
Annotation/
Real-Time
Clock
2.
Set
the
defibrillator/monitor
to
local
time.
(Press
TIME/DATE
MODE
on
the
rear
panel
to
display
the
clock
setting;
press
TIME/DATE
SET
to
increment
to
the
desired
value
in
the
displayed
field,
then
press
TIME/DATE
MODE
to
advance
to
the
next
field.)
3.
Press
RECORD
and
confirm
the
recorder
begins
printing
with
the
date
and
time
annotated.
4.
Confirm
that
the
time
is
accurate
to
within
+1
minute.
5.
Press
RECORD
to
stop
the
recorder.
PIP-Test
Load
1.
Make
sure
the
paddles
are
stored
in
the
paddle
storage
area
of
the
defibrillator/monitor.
2.
Press
W
ENERGY
SELECT
A
as
necessary
to
set
the
energy
level
to
2001.
3.
Press
CHARGE.
170008-000
A
3-10
All
Rights
Reserved
©
August
1993
Physio—Control

Table of Contents