LIFEPAK
9
defibrillator/monitor
3.
Press
Y
ECG
SIZE
A
to
scroll
to
the
DEFIB
test.
4.
Press
HR
ALARM
to
start
the
DEFIB
test.
5.
Press
Y
ENERGY
SELECTA
to
scroll
to
LOW.
Select
1J..
6.
Press
CHARGE.
7.
After
the
defibrillator/monitor
reaches
full
charge
as
indicated
by
the
tone,
simultaneously
press
both
paddle
discharge
pushbuttons
to
transfer
the
energy
into
the
energy
meter.
8.
Confirm
that
the
CRT
displays
the
message
JOULES
DELIVERED:
XXX,
where
XXX
is
the
energy
selected.
9.
Read
the
value
on
the
energy
meter.
10.
Press
Y
ECG
SIZE
&
as
necessary
to
change
the
JOULES
DELIVERED:
XXX
value
to
match
the
value
on
the
energy
meter.
11.
Press
HR
ALARM
to
enter
the
adjusted
value.
12.
Press
CHARGE.
After
charging
is
complete,
simultaneously
press
both
paddle
discharge
pushbuttons
to
transfer
the
energy
into
the
energy
meter.
Confirm
that
the
energy
meter
value
equals
the
selected
value
as
indicated
in
Table
3-8.
Table
3-8
Defibrillator
Calibration
Measurement
Ranges
Selected
Acceptable
Range
Selected
Acceptable
Range
Value
(J)
“Value
(J)
%
Tolerance
Value
(J)
“Value
(J)
%
Tolerance
1
0.8
to
1.2
+20%
20
19.0
to
21.0
+5%
2
1.8
to
2.2
+10%
30
28.5
to
31.5
+5%
3
2.7
to
3.3
+10%
50
47.5
to
52.5
+5%
4
3.8
to
4.2
+5%
100
95.0
to
105.0
+5%
5
4.8
to
5.2
+5%
200
190.0
to
210.0
+5%
6
5.7
to
6.3
+5%
300
285.0
to
315.0
+5%
7
6.7
to
7.3
+5%
360
342.0
to
378.0
+5%
8
7.6
to
8.4
+5%
9
8.6
to
9.4
+5%
10
9.5
to
10.5
+5%
13.
Press
HR
ALARM
to
reinitiate
the
DEFIB
test.
14.
Repeat
steps
5
through
13
for
each
energy
level.
15.
Return
the
paddles
to
the
paddle
storage
area
on
the
defibrillator.
x
170008-001
Orig
3-26
All
Rights
Reserved
©
August
1993
Physio—Control