ProMinent
®
Page 24
7.1.2 Menu overview: measured variable redox
NOTE
The following menu is only an example.
The displays can vary depending upon the general settings.
DMT settings
Redox
param. output signal
output: meas. value
4 mA: 0 mV
20 mA: 1000 mV
= Request access code
(if activated by the user)
mV
12.5 mA 21.0 ˚C
312
checking ORP-probe
probe in buffer!
faulty probe!!
buffer value: 465 mV
meas. value: 509 mV
check dry probe
resistance = 1.23 M
allowed > 2 M
faulty probe !!
general settings
DMTaW090R10E00220
2.1.06
general settings
language
DMTaW090R10E0220
E= english
param. temperature
input: sensor
unit: ˚C
device type
!! correct probe ??
type = ORP (Redox)
DMTaW090R10E0220
unit: °C
offset = 0.00 °C
meas. val. = 21.00 °C
output: meas. value
4 mA: 0 mV
20 mA: 1000 mV
general settings
change
access code = 5000
free access
BA_DM_163_03_08_GB.p65 26.03.2008, 10:20 Uhr24

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ProMinent DULCOMETER DMT Series and is the answer not in the manual?

ProMinent DULCOMETER DMT Series Specifications

General IconGeneral
BrandProMinent
ModelDULCOMETER DMT Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals