Page 44
9Troubleshooting
9.1 Error messages during operation
The following error messages can appear in the permanent display during
operation:
Message Displayed Cause of fault Remedy
value
Output overflow Measured Warning: detected current value Change configuration if
value greater than pre-set necessary
20 mA value
Output underflow Measured Warning: detected current value Change configuration if
value less than pre-set necessary
4 mA value
pH input error ! -2.00 pH Probe not present or lead Check probe connection
or damaged (switch off probe check if
16.00 pH necessary)
pH input error ! ~7.00 pH Short circuit Check probe connection
(probe resistance 0 )
defective pH probe !! Measured Measured probe Replace probe (switch off
value resistance < 2 M probe check if necessary)
ORP input error ! Measured value > 1200 mV or Check probe connection
< -1200 mV,
resp. fluctuating significantly
temp. input error ! 999.9 ° Probe not present or Check probe connection
lead damaged (set temperature measurement
to manual if necessary)
temp. input error ! -99.9 ° Short circuit Check probe connection
(set temperature measurement
to manual if necessary)
The error message disappears automatically once you have remedied the
fault.
Troubleshooting
BA_DM_163_03_08_GB.p65 26.03.2008, 10:20 Uhr44

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ProMinent DULCOMETER DMT Series and is the answer not in the manual?

ProMinent DULCOMETER DMT Series Specifications

General IconGeneral
BrandProMinent
ModelDULCOMETER DMT Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals