ProMinent
®
Page 31
DMT settings
pH Redox
general settings
DMTaW090P10E0220
2.1.06
device type
!! correct probe ??
type = pH
DMTaW090P10E0220
general settings
language
DMTaW090P10E0220
E= english
general settings
change
access code = 5000
free access
7.1.7 General settings
7.1.7.1 Changing the device type
You can select the measured variables, pH or redox potential, in the first
menu option “device type”.
Under “type” select pH or ORP (redox).
Depending upon your choice the corresponding identity code position alters:
pH = “P”, redox = “R”.
If you have confirmed the change of measured variable with the enter key
the output current will change immediately.
The settings for each measured variable are unaffected even after a different
selection has been made and are reinstated if the selection is reversed.
7.1.7.2 Changing general settings
You can adapt the DMT to your individual process requirements in the
second menu option “general settings”.
Access the individual identity code positions using the change key
.
Non-alterable features are automatically bypassed.
The following tables show the settings options depending upon the selected
measured value (the sequence from left to right corresponds to the identity
code position).
Measured variable pH
Language Automatic buffer Temperature Output Other
detection measurement settings
D = German 0 =ProMinent 0 = automatic 0 = measured variable; 0 =standard
E = English buffer set 1 = manual fault 23 mA,
F = French D = DIN 19266 2 = automatic or at calibration 23 mA
P = Polish buffer set manual 1 = manually adjustable
I = Italian V = automatic 9 = none current value
S = Spanish adjustment 2 = measured variable
or manual:
at fault 23 mA,
at calibration 23 mA
3 = measured variable
or manual:
at fault 23 mA,
when calibrating the last
measured value (HOLD)
4 = constant 4 mA
current
BA_DM_163_03_08_GB.p65 26.03.2008, 10:20 Uhr31

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ProMinent DULCOMETER DMT Series and is the answer not in the manual?

ProMinent DULCOMETER DMT Series Specifications

General IconGeneral
BrandProMinent
ModelDULCOMETER DMT Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals