DMT settings
Measured variable redox
Language Probe type Temperature Output Other
measurement settings
D = German 0 =standard 0 = automatic 0 = measured variable: 0 =standard
E = English 9 = none at fault 23 mA,
F =French at calibration 23 mA
P = Polish 1 = manually adjustable
I = Italian current value
S = Spanish 2 = measured variable
or manual:
at fault 23 mA,
at calibration 23 mA
3 = measured variable
or manual:
at fault 23 mA,
when calibrating the last
measured value (HOLD)
4 = constant 4 mA
current
7.1.7.3 Changing access code
The factory pre-set access code is 5000. This setting allows unlimited access
to the calibration menu and the menu options in the settings level.
This code can be changed in the third menu option “change access code”.
Use the change key
to move to individual digits.
The new code is confirmed and activated with the enter key .
NOTE
• If an access code has been activated, both the calibration
menu and the menu options are access-protected.
To gain access, enter the access code when requested
to do so and confirm using the enter key
. Access
protection is reactivated when the DMT returns to
continuous display mode.
• Deactivate the access code by resetting the code to 5000
in the “change access code” menu option.