ProMinent
®
Page 9
2 Safety
2.1 Correct use
The DMT on-site measurement transducer is designed exclusively for the
• measurement of pH value resp. redox potential
• measurement of temperature
• display of measured variables
•production of an output signal
It is prohibited to use the device for any other applications or to modify it
in any way!
The device must not be used outdoors without added protection (housing,
weatherproof cover).
2.2 Safety guidelines
CAUTION
• The device must not be used in a possible explosion
area.
• The DMT must be operated by trained and authorised
personnel.
IMPORTANT
• The system must be suitably equipped and configured to
prevent overdosing of hazardous materials due to probe
failure.
• The DMT has no on/off switch. It starts to function as soon
as it is connected to a power supply.
3 Storage and transport
Store and transport the DMT in the original packaging.
IMPORTANT
• Protect the DMT from damp and the effects of chemicals
even when packed.
Ambient conditions for storage and transport:
Temperature: -20 °C to 70 °C
Humidity: max. 95 % relative humidity,
non condensing
Safety / Storage and transport
BA_DM_163_03_08_GB.p65 26.03.2008, 10:20 Uhr9