8.2.2 The calibration process
NOTE
The back key can be used to exit the calibration menu at
any time. You will then return to the permanent display.
The output current is frozen for the duration of the calibra-
tion at, depending upon general settings, the last
measured value (HOLD) or at 23 mA (see chapter 7.1.7
General settings).
Inadmissible values render the ongoing calibration invalid.
The previous calibration values are retained. The error
messages are explained in chapter 9.2.
Please dispose of spent buffer solution.
Calibration menu overview
8.2.2.1 2 point calibration
Operating the DMT
BA_DM_163_03_08_GB.p65 26.03.2008, 10:20 Uhr40