EasyManua.ls Logo

ProMinent DULCOMETER DMT Series - Page 41

ProMinent DULCOMETER DMT Series
56 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ProMinent
®
Page 41
8.2.2.2 1 point calibration
Proceed as follows to call up the calibration menu:
pH
7.12
input: manual
buffer temp. = 20.3 ˚C
HOLD 7.12 pH
HOLD 12.12 mA
input: sensor
buffer temp. = 20.3 ˚C
HOLD 7.12 pH
HOLD 12.12 mA
pH
12.5 mA 21.0 ˚C
or
Starting
calibration Press the CAL key when in the permanent display mode while the pH
probe is immersed in the liquid sample.
Under “input” set the detection method of the buffer temperature
(correction value):
Sensor (if the pH sensor incorporates a temperature gauge) or
manual (if using a separate temperature gauge).
If you select manual, you must set the buffer temperatures under “buffer
temp.” using the up and down arrow
/ keys.
This setting is only valid during calibration.
You can change the pH HOLD value under “HOLD”. This setting is only
valid during calibration.
Remove the pH probe from the liquid sample, rinse and immerse in the
first buffer solution.
Press the CAL key to start automatic buffer detection:
Operating the DMT
BA_DM_163_03_08_GB.p65 26.03.2008, 10:20 Uhr41

Table of Contents

Related product manuals