7.2.4 General settings
T
general settings
DMTaW090T10E0020
2.1.06
device type
!! correct probe ??
type = temp.
DMTaW090T10E0020
general settings
language
DMTaW090T10E0020
E = english
general settings
change
access code = 5000
free access
7.2.4.1 Changing device type
The first menu option, “device type”, is used to switch between the measured
variables pH value, redox voltage or temperature:
Under “type”, select pH, ORP (redox) or Temperature.
The corresponding identity code position changes depending on your choice:
pH= “P”, redox = “R”, temperature = “T”.
If you confirm the change of measured variable using the enter key
,
the output current changes immediately.
The settings for each measured variable are unaffected by switching from one
to another, i.e. you do not need to re-set them when re-selecting a measured
variable.
7.2.4.2 Changing general settings
The second menu option; “general settings”, is used to adapt the DMT to
your specific process requirements.
Scroll through individual identity code positions using the “Change” key
until you reach the one you require. Non-adjustable features are automatically
bypassed.
The following tables show the settings options for each measured variable
(the sequence from left to right corresponds to the identity code positions).
Measured variable temperature
Language Probe type Temperature Output current Additional
input settings
D = German standard 0 = sensor 0 = measured variable; 0 =standard
E = English 0 =PT 100 or PT 1000 (100) at fault 23 mA
F = French PT 1000 at calibration 23 mA
P = Polish 1 = adjustable output
I = Italian 2 = measured variable
S = Spanish or adjustable;
at fault 23 mA
at calibration 23 mA
4 =4 mA constant current
DMT settings
BA_DM_163_03_08_GB.p65 26.03.2008, 10:20 Uhr36

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ProMinent DULCOMETER DMT Series and is the answer not in the manual?

ProMinent DULCOMETER DMT Series Specifications

General IconGeneral
BrandProMinent
ModelDULCOMETER DMT Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals