7.1.5.2 General setting: Temperature measurement = manual
(pH only)
pH
T = 22.6 ˚C
unit: ˚C
param. temperature
input: manual
unit: ˚C
Enter the process temperature under “T”.
Select the temperature unit °C or °F under “unit”.
7.1.5.3 General setting: Temperature measurement = automatic or
manual (pH only)
pH
input: sensor
unit: ˚C
offset = 0.00 ˚C
meas. val. = 21.00 ˚C
or
param. temperature
input: manual
unit: ˚C
input: manual
T = 22.6 ˚C
Unit: ˚C
Under “input” select sensor or manual:
Select sensor if a temperature gauge is connected (temperature
measurement = automatic):
Select manual if the user is going to enter the process temperature
(temperature measurement = manual).
The selection made under “input” affects subsequent settings options.
Sensor If you have selected sensor in the “input” option:
Select the unit of measurement for temperature °C or °F under “unit”.
Enter the difference t under “offset” to a reference temperature
measurement (see 7.1.5.5 Calibrating the Pt 100 temperature gauge).
The actual temperature measured is displayed under “meas. val.”.
Manual If you have selected manual in the “input” option:
Enter the process temperature under “T”.
Select the temperature unit °C or °F under “unit”.
7.1.5.4 General setting: Temperature measurement = none
No info display appears in the operating menu.
No temperature measurement is carried out.
DMT settings
BA_DM_163_03_08_GB.p65 26.03.2008, 10:20 Uhr27