18 65000159 Rev. 008 (06/17)
Calibration Procedure
Action
Instrument
Information
Illustration
Place the printed
Calibration label on a Label
Record in the appropriate
assay section of your
Quality Assurance Log.
Important calibration
information is now
included in your Quality
Assurance Log.
Discard the FastPack
®
IP
into a Biohazards
container.
Repeat the Calibration
Procedure.
All calibrations require that
TWO (2) calibrations be
run with the calibrator.
The exception is the
FastPack® IP Vitamin D
Immunoassay which
requires THREE (3)
calibrations. All
calibrations must “PASS”
for a successful calibration.
See the Calibration Process
Tree on Page 20 which
illustrates the messages
shown on the analyzer
display.
See Direction Insert for the
test being calibrated.
Discard all expired
calibrator material, in
stoppered vials, into a
Biohazard container.
Order a new Calibration Kit
from your local distributor
or Qualigen, Inc.