26 65000159 Rev. 008 (06/17)
Control Testing Procedure
Remove and inspect the
FastPack
®
IP. Check the
inside of the Analyzer door
for fluid. If there is fluid on
the FastPack
®
IP or
Analyzer, disregard the
result and properly dispose
of the pack. Implement the
cleaning procedure.
Ensure that all chambers
have burst. Fluid should
easily flow between all
chambers. Observe that
most of the fluid in the
FastPack
®
IP is yellow. If
there is no yellow fluid,
disregard the result and
rerun FastPack
®
IP.
Slip the FastPack
®
IP into
the external printer from
the left side with the label
side up and the label facing
towards the rear of the
external printer.
Printer will print and
release the FastPack
®
IP
automatically.
It is important to insert the
FastPack
®
IP from the left
side of the printer to avoid
tangling or twisting the
printer ribbon.
Remove the FastPack
®
IP
from the left side of the
external printer.
It is important to remove
the FastPack
®
IP from the
left side of the printer to
avoid tangling or twisting
the printer ribbon.
Find the result printed on
the FastPack
®
IP peel-off
label.
Control Result
Units
Date of Test
Instrument Serial#
All the necessary
information for Quality
Control is now included on
the peel-off label.
FastPack
®
IP Label after Printing
FastPack IP Expiration
Lot Number Date
Control Date of Instrument Operator
Result Test Serial # Initials