65000159 Rev. 008 (06/17) 27
Control Testing Procedure
Action
Instrument
Information
Illustration
Match each control lot
number from the Control
Range Card with the lot
number on the control
vials.
If each of these lot
numbers match, then
verify that the result
printed on the FastPack
®
IP
is within range for the
appropriate control.
If a control is outside of
the acceptable range,
repeat the test. If the
retest is still outside of the
acceptable range, call
Qualigen System Support
at 877-770-6127.
Apply the Control Label
onto the Label Record in
the appropriate assay
section of your Quality
Assurance Log.
Important control
information is now
included in your Quality
Assurance Log.
Discard the used FastPack
®
IP into a Biohazard
container.
The FastPack
®
IP analyzer
requires testing with two
levels of controls. Repeat
the control procedure as
needed.
Discard both control vials
when one vial is empty or
when the control kit has
expired into a Biohazard
container.
Order a new control kit
from your local distributor
or Qualigen, Inc.
Match Control
Lot Numbers
Result
must be
within
range.