52 65000159 Rev. 008 (06/17)
FastPack
®
IP System Procedure
Action
Instrument
Information
Illustration
Remove and inspect the
FastPack
®
IP. Check the
inside of the analyzer door
for fluid. If there is fluid on
the FastPack
®
IP or the
analyzer, disregard the
result and properly dispose
of the pack. Implement the
cleaning procedure.
Confirm that all chambers
have burst. Fluid should
easily flow between all
chambers. Observe that
most of the fluid in the
FastPack
®
IP is yellow. If
there is no yellow fluid,
disregard the result and
rerun a new FastPack
®
IP.
Slip the FastPack
®
IP into
the external printer from
the LEFT SIDE with the
label side up and the label
facing towards the rear of
the external printer. Be
sure that the edge of the
FastPack
®
IP goes to the far
edge of the external
printer.
The printer will print and
release the FastPack
®
IP
automatically. It is
important to insert the
FastPack
®
from the left side
of the printer to avoid
tangling or twisting the
printer ribbon.
Remove the FastPack
®
IP
from the LEFT SIDE of the
external printer.
It is important to remove
the FastPack
®
IP from the
left side of the printer to
avoid tangling or twisting
the printer ribbon.
Find the result printed on
the FastPack
®
IP peel-off
label.
Patient Result
Units
Date of Test
Instrument Serial #
All the necessary
information for Quality
Control is now included on
the peel-off label.
FastPack
®
IP label after Printing
Result FastPack IP Expiration Instrument
Lot Number Date Serial
Number
Patient Date Test Operator
Name Type Initials