Communication Protocol CubiScan 125 Command Set
CubiScan 125 Service Manual 70
* This field may contain underscores, dashes, or tildes indicating an under, unstable, or over
error condition, respectively. Leading spaces (20h) will be used when the actual data does not fill
the entire field.
Scale Calibration
This function is required when the load cell is replaced, when a new control is installed, or for
routine calibration. This command causes the CubiScan 125 to reply with a scale calibration code.
This command must be issued THREE times to complete scale calibration. Scale calibration
codes are defined below.
7 1 Carriage Return Control (CR) 0Dh
81Line Feed Control(LF) 0Ah
Pos Len Description Type Range ASCII
Command Format
1 1 Start of Text Control (STX) 02h
2 1 Scale Calibration Alpha (S) 53h
3 6 Test Weight Value Numeric 050.00-100.00
91End of Text Control(ETX) 03h
10 1 Carriage Return Control (CR) 0Dh
11 1 Line Feed Control (LF) 0Ah
Acknowledge Format
1 1 Start of Text Control (STX) 02h
2 1 Scale Calibration Alpha (S) 53h
3 1 Acknowledge Alpha (A) 41h
4 2 Identifier Numeric (00-99)
61End of Text Control(ETX) 03h
7 1 Carriage Return Control (CR) 0Dh
81Line Feed Control(LF) 0Ah
Negative Acknowledge Format
1 1 Start of Text Control (STX) 02h
2 1 Scale Calibration Alpha (S) 53h
31Neg. Acknowledge Alpha (N) 4Eh
41End of Text Control(ETX) 03h
5 1 Carriage Return Control (CR) 0Dh
61Line Feed Control(LF) 0Ah
Pos Len Description Type Range ASCII