clone Portable Gamma-Ray Spectrometer, Revision 10 1.0 GENERAL SYSTEM DESCRIPTION
Ra
dComm – Proprietary Page 6 Revision Date: March 3, 2016
h)
BATTERIES - The Syclone requires manufacturers’ supplied lithium batteries.
While rechargeable batteries are shipped fully charged, they will self-discharge with time
when not in use. It is important to condition new batteries, whether received with the system
or purchased separately. Conditioning is achieved by first totally discharging the batteries in
the SYCLONE unit and then fully charging them for at least 4 hours.
Note: To get reliable performance and full capacity out of the batteries throughout their life, it
is important to let them fully discharge and then fully re-charge for 4 hours at least
once every two months. This will assure that the batteries provide up to 12 hrs(typical
9 hours) of operation.
i) BATTERY-CHARGER – the battery charger is an integral part of the unit and requires the
supplied power supply, which plugs into a normal wall receptacle using house current.
j) CHECK SOURCE – The Syclone uses a 0.25 µCi Cs-137 check source, which will stabilize
the unit under 3 mins.
Note: There is a central mark on the front of the case which represents the effective center
of the NaI detector. Place the supplied source at this spot when the unit requests
stabilization, and follow the screen messages.
k) SYSTEM SUPPORT SOFTWARE - The RADVIEW software is provided with the Syclone
unit on a USB. This is a Windows based program that operates under Windows Vista,
windows 7 and 8. It provides data downloading, data display, custom library uploading,
spectrum display analysis and various other features.
l) HARD SHELL TRANSPORTATION CASE - The system is supplied with a hard shell
carrying case for easy transportation of the Syclone.
m) USER MANUAL - This user manual is supplied with the instrument.
To charge the unit and connect to
the
PC to transfer data using the mini
USB cable.