- 122 -
7. On inserting the data, the program will pass on to the proper process of
measuring.
8. First, the sinker should be placed on the hook (a sample mass determination in
the air), and after the indication stabilization the measurement result is
approved
9. Next, take the sinker out of the hook and put the tested sample with liquid on
the beaker’s basis (the beaker cannot touch the hook). Delicately hang the
sinker on the hook (the sinker must be totally emerged in the tested liquid) a
mass sample determination in liquid and after the indication stabilization
approve the result.