- 24 -
3. Remove the side glass walls. The side glass walls are not replaceable,
so it is important to remember which one is left and right, and after
cleaning install them back properly in the draft shield.
4. CAUTION: do not remove the front glass wall.
5. Remove the weighing pan, the weighing pan safety rings and the
bottom metal plates, so as not to damage the balance mechanism durin
the cleaning.
Thus prepared draft shield and glass walls can be properly cleaned. All the
operations should be done carefully. Pay special CAUTION to the place
where the weighing pan was installed: dirt and other small elements might
enter the balance construction through this openiing, which might negatively
influence the balance parametres.
Upon completion of the cleaning proces reassembly the balance. To do it
repeat steps 1-5 in a reverse order. Be especially careful while assembling
side glass walls, they must take the right place. While sliding the glass
back onto its place remember to pull the protection (like before
when dismantling the draft shield).