Trying Out the Instrument
R&S
®
SMW200A
57Getting Started 1412.9220.02 ─ 21
To reconfigure the trigger settings
We assume that the instrument is configured as described in Chapter 6.2, "Gen-
erating a Digitally Modulated Signal", on page 51 and the default connector map-
ping is maintained (see Figure 6-5).
1. In the block diagram, select "Baseband > Misc > Custom Digital Mod > Trigger
In".
2. Select the following settings:
a) "Mode > Armed Auto"
b) "Source > External Global Trigger 1".
The instrument expects an external global trigger event. In this current config-
uration (see Figure 6-5), the "Global Trigger 1" signal has to be supplied at the
input connector USER 3.
The Trigger/Marker/Clock status LEDs in the block diagram confirm that an
external trigger signal is selected; the signal generation is however stopped.
Triggering the Instrument with an External Signal