R&S ZVA Repairs
1177.5333.02 3.91 E-11
Group 1:
General Functions
Functions
Service
function
Data
Serv. Lev.
Enables/disables the peak detector and RMS
detector independently of the model
1.0.0.1.X
0 Use disabled
1 Use enabled
2
Enable/disables the arbitrary mode settings in
the port configuration (e.g. LowNoise/
LowDistortion) independently of the model
1.0.0.5.X
0 Use disabled
1 Use enabled
2
Activates or deactivates the setting of measured
values to default values when status messages
are issued due to HW faults
1.1.0.2.X
0 Deactivate default values
1 Activate default values
0
Activates or deactivates factory calibration for
the active setup
1.1.0.3.X
0 Deactivate factory calibration
1 Activate factory calibration
0
Group 2:
General Functions
Functions
Service function
Data
Serv. Lev.
Suppress error-message box
2.0.0.0.X
X=0 MSG box is output (default state)
X=1 MSG box is suppressed (error is
nevertheless entered in the log file)
1
Selftest all boards
2.0.0.5.0.0
0, 1 or 2
Network controller
Functions
Service function
Data
Serv. lev.
ND1:Read Temp (addr.:68)
2.1.2.3
1
ND2:Read Temp (addr.:68)
2.2.2.3
1
ND[1..2] Selftest
2.[1-2].2.5.0.0
0, 1 or 2
NA1:Read Temp (addr.:69)
2.1.3.3
1
NA2:Read Temp (addr.:69)
2.2.3.3
1