R&S ZVA Repairs
1177.5333.02 3.113 E-11
Screw the OPEN from the calibration kit to the port connector.
Select User Def Math : (lin Mag(Data) + (linMag(Mem2)) / 2 , activate Result is Wave Quantity
Select Math = User Def
With the stated levels, the cable loss must still be taken into account.
If the measured values are more than 2 dB below the levels given in the table, the coupler unit must be
replaced.