Repairs R&S ZVA
1177.5333.02 3.118 E-11
With the stated levels, the cable loss must still be taken into account.
If the measured values are more than 2 dB below the levels given in the table, the coupler unit must be
replaced.
Coupler Directivity
The following is assumed:
The generator and receiver sections of the reflectometer associated with the coupler are OK.
Perform a sweep from 10 MHz to 67 GHz, measure Ratio bx/ax Source Portx
Screw the SHORT from the calibration kit to the port connector, save measured values (Data ->
Mem).
Screw the OPEN from the calibration kit to the port connector, save measured values (More Mem :
Data -> Mem3).
Screw the MATCH from the calibration kit to the port connector.
Select User Def Math : linMag(Data) / (linMag(Mem2) + linMag(Mem3)) / 2 : Math = User Def
The trace gives the directivity.
If the measured values are more than 2 dB greater than the values stated in the table, the coupler must
be replaced.