R&S ZVA Repairs
1177.5333.02 3.121 E-11
Enter service function 2.21.2.18.17512345 (IF = 17.512345 MHz).
Disable level corrections with SF 2.21[..24].2.15.1
If the level displayed on the ZVA's screen is not within the range -4 dBm ± 2 dB, the board must be
replaced.
Testing the Motherboard
28 V supply
Using a multimeter, measure the voltage at X 100.B5 (wrt ground).
Permissible deviation: ± 0.5 V
Preamplifier for DC measurement inputs
Apply the DC voltages listed in the table using the 4-pin Mini-DIN connector at the DC measurement
input.
Measure the DC voltage with a multimeter.
Fig. 3-11 Pin assignment DC MEAS connector
If the measured value is more than 10% above or more than 10% below the stated nominal value, the
motherboard must be replaced.