Descaling the humidifying nozzle
Please check the humidifying nozzle for limescale
deposits once a month.
Caution!
• The process described involves working
with an aggressive chemical fluid – danger
of corrosive burns!
• To prevent damage to health
it is vital that you use:
protective clothing
safety goggles
protective gloves
face mask
1
Switch unit off.
2 To descale the humidifying nozzle please
remove the left grid shelf.
3 Open the air baffle.
4 Release the union nut of the humidifying nozzle
(turn anticlockwise – spanner size 18).
5 Place the humidifying nozzle in the descaling
fluid overnight. Before assembling make sure
that the hole in the humidifying nozzle is free of
limescale.
6 Assemble in reverse order. Take care to fit the
humidifying nozzle in the correct position.
Note:
Make sure that the air baffle and the grid shelf are
properly fitted and locked.
You will need: descaler (10 litre canister – Part
number 6006. 0110).
- 56 -
Service – Humidifying nozzle descaling