4.5.13 Motor Test Run of the Frequency Module Page 58
4.5.13 Motor Test Run of the Frequency Module
Setup Number: 50
The user can test the mechanically driven frequency module of the Rayocomp
PS 10 with this setup function. If your Rayocomp has been subject to jostling
and vibration, perhaps during long transportation, you can make sure that
the frequency module is still in perfect working order with this function.
After you have entered the setup number 50 and confirmed your entry with
the OK-key, the Rayocomp PS 10 starts a trial run. You can read the progress
of the trial run from your display. At the end, the device should display the
following message:
The 1 displayed behind the 50 indicates an optimal functioning of the
frequency module. If you see a 0 in the same place, however, the trial run
was faulty. Please repeat the trial run in this case. If the 0 appears again, your
Rayocomp PS 10 requires servicing and has to be sent back to us.