Troubleshooting
English 71
Error code Action
504 Unable to learn the circuit.
1. Ensure that the circuit is not moved until completion of the test.
2. Check the circuit and attached accessories for blockages.
3. Ensure that there are no sharp bends or kinks in the circuit and the patient end is not
blocked.
4. If using humidification, ensure that the humidifier tub is not overfilled.
5. Follow the on-screen instructions carefully:
circuit should not be blocked during step 2
circuit should be completely blocked during step 3.
If the problem persists, contact your clinician or an authorised Service Centre.
505 The measured circuit resistance exceeds safe operating limits for this device.
1. Check the circuit and attached accessories for blockages.
2. Ensure that there are no sharp bends or kinks in the circuit and the patient end is not
blocked.
3. If using humidification, ensure that the humidifier tub is not overfilled.
4. Follow the on-screen instructions carefully:
circuit should not be blocked during step 2
circuit should be completely blocked during step 3.
If the problem persists, contact your clinician or an authorised Service Centre.
506, 512 Hardware fault detected. Contact an authorised Service Centre.
600 Unable to calibrate Expiratory Flow Sensor.
1. Check the circuit adapter is clean, dry and firmly inserted.
If the adapter is wet, then removing the adapter and vigorously shaking to clear
water can be effective. Re-insert adapter firmly and repeat Learn Circuit.
If the adapter is not clean, then it will need to be replaced.
2. If using a small diameter paediatric breathing circuit, consider using a bacterial/viral
filter or a 22-mm adapter on the expiratory adapter port.
3. Repeat Learn Circuit and ensure that the circuit is not moved until completion of the
test.
If the problem persists, contact an authorised Service Centre.

Table of Contents

Other manuals for ResMed Astral Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ResMed Astral Series and is the answer not in the manual?

ResMed Astral Series Specifications

General IconGeneral
TypeVentilator
Displaycolor LCD
ConnectivityUSB
EPAP0 to 25 cm H2O
Rise TimeAdjustable
PEEP0 to 25 cm H2O
FiO2 MonitoringYes, with optional oxygen module
Battery BackupInternal and external battery options
Oxygen CapabilityYes, with optional oxygen module
AlarmsLow battery
Pressure Support0 to 30 cm H2O

Related product manuals