Technical specifications
74
Data storage
7 days of high-resolution airway pressure, respiratory flow and delivered
volume (sampled at 25 Hz).
7 days of breath-related therapy data (sampled at 1 Hz).
365 days of statistical data per program.
Dimensions (L x W x H)
285 mm x 215 mm x 93 mm
Display screen size: 150 mm x 90 mm
Weight
3.2 kg
Inspiratory port / double limb adapter
22 mm taper, compatible with ISO 5356-1
Anaesthetic & Respiratory Equipment Conical Connectors
Pressure measurement
Internally mounted pressure transducers
Flow measurement
Internally mounted flow transducers
Power supply
AC (100-240V), (50-60Hz), 90 W 3.75 A continuous, 120 W / 5A peak
External DC Power Supply
12 - 24V DC 90 W, 7.5 A / 3.75 A
Internal Battery
Lithium-Ion battery, 14.4 V, 6.6 Ah, 95 Wh
Operating hours (standard case): 8 hours with a new battery under normal
conditions.
Test conditions: Adult, (A)CV mode, Vt = 800 mL, PEEP = 5 cmH
2
O, Rate = 20
bpm, Ti = 1.0 sec. All other parameters remain at default settings.
Test lung configuration: R = 5 hPa (L/s)-1, C = 50 mL (hPa)-1
Operating hours (worst case) > 4 hour run time under the following
conditions:
Test conditions: Adult, non-vented, PACV mode, Double limb circuit,
Pressure Assist = 30 cmH
2
O, PEEP = 20 cmH
2
O Rate:20 bpm,
Ti: 1.0 sec, Rise Time = Min, Safety Vt = Off, Trig = Off. All other
parameters remain at default settings.
Total lifetime: up to 3,000 hours of normal operation on internal battery
Note: Time may vary with different settings and environmental conditions.
Housing construction
Flame retardant engineering thermoplastic
Environmental conditions
Operating temperature: 0°C to 40°C
Charging temperature: 5°C to 35°C
Operating humidity: 5 to 93% non-condensing
Storage and transport temperature: -25°C to 70°C for up to 24 hours
Storage and transport temperature: -20°C to 50°C for greater than 24 hours
Note: Storing the Astral device at temperatures exceeding 50°C for
extended period of time may accelerate battery aging. This will not affect
the safety of the battery or device. Refer to Using the internal battery (see
page 46)
Storage and transport humidity: 5 to 93% non-condensing
It takes 40 minutes* for the device to be ready for use on a patient when
removed from storage at the minimum long term temperature and at an
ambient temperature of 20°C.
*Assumes that the device is connected to an external AC power.
It takes 60 minutes for the device to be ready for use on a patient when
removed from storage at the maximum long term temperature and at an
ambient temperature of 20°C.

Table of Contents

Other manuals for ResMed Astral Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ResMed Astral Series and is the answer not in the manual?

ResMed Astral Series Specifications

General IconGeneral
TypeVentilator
Displaycolor LCD
ConnectivityUSB
EPAP0 to 25 cm H2O
Rise TimeAdjustable
PEEP0 to 25 cm H2O
FiO2 MonitoringYes, with optional oxygen module
Battery BackupInternal and external battery options
Oxygen CapabilityYes, with optional oxygen module
AlarmsLow battery
Pressure Support0 to 30 cm H2O