User Manual
3
• If you notice any unexplained changes in the performance of the device, if it is making unusual
or harsh sounds, if it and/or the power supply has been dropped or mishandled, if the enclosure is
broken, or if water has entered the device, discontinue use and contact your home care provider.
• Repairs and adjustments must be performed by Respironics - authorized service personnel only.
Unauthorized service could cause injury, invalidate the warranty, or result in costly damage.
• Periodically inspect electrical cords, cables, and the power supply device for damage or signs of
wear.
• To avoid electrical shock, unplug the device before cleaning.
• Pins of connectors identified with the ESD warning symbol should not be touched. Connections
should not be made to these connectors unless ESD precautionary procedures are used.
Precautionary procedures include methods to prevent build-up of electrostatic discharge (e.g., air
conditioning, humidification, conductive floor coverings, non-synthetic clothing), discharging
one’s body to the frame of the equipment or system or to earth or a large metal object, and
grounding oneself by means of a wrist strap to the equipment or system or to earth.
2.2 Cautions
• e device may only be operated at temperatures between 41° F (5° C) and 95° F (35° C).
• A properly installed, undamaged reusable foam inlet filter is required for proper operation.
• Do not immerse the device or allow any liquid to enter the enclosure or the inlet filter.
• Condensation may damage the device. Always allow the device to reach room temperature before
use.
NOTE: Additional warnings, cautions, and notes are located throughout this manual.
2.3 intended use
e BiPAP AVAPS device is intended to provide noninvasive ventilation for pediatric patients 7 years or
older > 40 lbs (18.2 kg) and adult patients > 66 lbs (30 kg) with respiratory insufficiency or obstructive
sleep apnea. is device may be used in the hospital or home.
NOTE: e device is intended for use with nasal masks and full-face masks as recommended by
Respironics.
NOTE: e device is to be used only on the instruction of a trained health care professional.
WARNING: e effectiveness of Bi-Flex, C-Flex or AVAPS therapy has not been established for
pediatric patients at this time.